~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
| Unit 29
Spectroscopy & Structure
Techniques & Instrumentation
|

|
~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
Overview
One
primary consideration in selecting an analytical technique for identification
purposes is the need for either a qualitative or a quantitative
determination. The former relates just to the identity of the material, while
the latter requires the precise determination of amounts, quantities, and
percent compositions by weight.
Now that
the basic components of matter have been defined, the proper selection of
analytical techniques that will allow the analytical chemist to identify or
compare matter can best be understood by classifying all substances into one of
two very broad chemical groups: organics and inorganics. Organic
substances are based the element carbon, and comprise all forms of matter and
energy which live and breathe. Inorganic substances encompass all other known
chemical substances. Each of these two broad categories has both chemical and
physical properties that are quite distinctive and characteristic.
Organic compounds consist of carbon atoms, hydrogen atoms, and functional
groups. The valence of carbon is 4, and hydrogen is 1, functional groups are
generally 1. From the number of carbon atoms and hydrogen atoms in a molecule
the degree of unsaturation can be obtained. Many, but not all structures can be
envisioned by the simple valence rule that there will be one bond for each
valence number. The knowledge of the chemical formula for an organic compound is
not sufficient information because many isomers (different structural forms with
the same chemical formula) can exist.
Techniques of elemental analysis are
generally destructive in their use for the determination of elemental
composition of individual compounds and molecules. Spectroscopy is the method
most often used, it all its various forms. The instruments most directly
concerned with our primary interest, molecular structure, are the
spectrometers - measurers of spectra. Of the various spectra, we shall work
chiefly with only two kinds: infrared (IR) and nuclear magnetic
resonance (proton NMR and CMR), since they are the workhorses of the modern
organic chemistry laboratory. Of these, we shall devote an entire chapter to
NMR. We shall look briefly at three other kinds of spectra: mass, ultraviolet
(UV) and electron spin resonance (ESR).
Organic compounds often exist as mixtures. Because many organic compounds
have relatively low boiling points and/or dissolve easily in organic solvents
there exist many methods for separating mixtures into pure constituents that are
specific to organic chemistry such as distillation, crystallization and
chromatography techniques. Thus, there exist a number of different methods for
deducing the structure and/or chemical composition an organic compound. These
include:
IR Spectroscopy, Atomic Spectroscopy, Mass
Spectrometry, Nuclear Magnetic Resonance (NMR), X-Ray Diffraction, UV/Vis
Spectroscopy, and Chromatography, all of which will be briefly
described in this unit.
Spectroscopy
Spectroscopy is the study
of matter and its properties by investigating light, sound, or
particles that are emitted, absorbed or scattered by the matter
under investigation.
Spectroscopy
may also be defined as the study of the interaction
between light and matter. Historically, spectroscopy
referred to a branch of science in which visible light
was used for theoretical studies on the structure of matter and
for qualitative and quantitative analyses. The definition has broadened,
however, as new techniques have been
developed that utilize not only visible light, but many
other forms of electromagnetic and non-electromagnetic radiation,
including
microwaves,
radiowaves,
x-rays,
electrons,
phonons (sound waves) and others.

Impedance spectroscopy
is a study of frequency response in alternating current.
Spectroscopy is often used for the identification of
substances through the frequency spectrum emitted or absorbed by
them. A device for recording a spectrum is a
spectrometer. Spectroscopy can be classified according to
the physical quantity which is measured or calculated or the
measurement process.
I. IR
Spectroscopy
Of all the
properties of an organic compound, the one that gives the most
information about the compound's structure is its infrared
spectrum. Thus,
infrared spectroscopy is used to determine the presence of
certain functional groups. Infrared frequencies of light
are used to study fundamental vibrations and associated
rotational-vibrational structure via vibrational resonance
and selective absorption.
Infrared
spectroscopy works because chemical bonds have
specific frequencies at which they vibrate
corresponding to energy levels. In order for a
vibrational mode in a molecule to be IR active,
it must be associated with changes in the
permanent dipole. The resonant frequencies are
determined by the
normal modes of vibration. (Normal modes are
also known as natural frequencies or resonant
frequencies. There is a set of these frequencies
that are unique to each molecular structure).
These natural frequencies can be related
to the strength of the bond, and the mass of the
atoms (or harmonic oscillators) at
either end of it. Thus, the frequency of the
vibrations can be associated with a particular
bond type.
Changes in the
vibrational frequencies of a molecule are caused
by the absorption of infrared light. A
particular portion of the infrared spectrum is
referred to either by its wavelength (in microns
or Angstroms) of frequency (not Hz, but rather
in wavenumbers, or reciprocal centimeters. The
wavenumber is simply the number of wave periods
per centimeter, and is equal to the reciprocal
of the wavelength in centimeters.
A particular
group of atoms gives rise to characteristic
absorption bands. For example, the
OH group absorbs strongly at 3200-3600 per
cm; the C=O group of ketones absorbs
strongly at 1710 per cm; the C=N group
absorbs strongly at 2250 per cm; and the CH3
group absorbs at both at 1450 and 1375 per
cm (2 different modes of vibration).
Interpretation
of an IR spectrum is not a simple matter. Bands
may be obscured by the overlapping of other
bands. Overtones (or harmonics) may appear at
exactly twice the frequency of the fundamental
band. The absorption band of a particular group
may be shifted by various structural features,
such as conjugation, electron withdrawal by a
neighboring substituent, angle strain or van der
Waals strain, or hydrogen bonding.
Characteristic IR Absorption Frequencies

Simple diatomic
molecules have only one bond, which may stretch.
More complex molecules may have many bonds, and
vibrations can be conjugated, leading to
infrared absorptions at characteristic
frequencies that may be related to chemical
groups. The atoms in a CH2 group,
commonly found in organic compounds can vibrate
in
six different ways, symmetrical and
asymmetrical stretching, scissoring,
rocking, wagging and twisting.
In
order to measure a sample, a
beam of infrared light is passed
through the sample, and the
amount of energy absorbed at
each wavelength is recorded.
This may be done by scanning
through the spectrum with a
monochromatic beam, which
changes in wavelength over time,
or by using a Fourier transform
(FTIR) instrument to measure all
wavelengths at once. From this,
a transmittance or absorbance
spectrum may be plotted, which
shows at which wavelengths the
sample absorbs the IR, and
allows an interpretation of
which bonds are present.
The infrared portion of the
electromagnetic spectrum is
divided into three regions; the
near-IR, mid-IR and far-IR
(relative to the visible
spectrum).
1) The far-IR
(~ 400-10 cm-1).
Adjacent to the
microwave
region, the
longer
wavelength
far-IR has lower
energy and may
be used for
rotational
spectroscopy.
2) The mid-IR
(~ 4000-400 cm-1)
may be used to
study the
fundamental
vibrations and
associated
rotational-vibrational
structure.
3) The
near-IR
(14000-4000 cm-1)
with its higher
energy level and
higher
frequencies can
excite
overtone or
harmonic
vibrations.
This technique works almost
exclusively on covalent bonds,
which are those found
predominantly in organic
compounds. Clear spectra are
obtained from samples with few
IR active bonds and high levels
of purity. More complex
molecular structures lead to
more absorption bands and more
complex spectra. The technique
has been used for the
characterization of very complex
mixtures.
II. Atomic Spectroscopy
Atomic spectroscopy,
such as flame atomic absorption, graphite furnace atomic absorption, and
inductively coupled plasma atomic emission, are used to probe the
outer electronic structure of atoms. Other spectroscopic methods which
are used to probe the inner electronic structure of atoms include
X-ray fluorescence,
particle induced x-ray emission,
x-ray photoelectron spectroscopy, and
Auger electron spectroscopy.
Atomic spectroscopy is the
determination of elemental composition either by its
electromagnetic spectrum or its mass spectrum. Atomic
spectroscopy is closely related to other forms of spectroscopy.
It can be divided by atomization source or by the type of
spectroscopy used. In the latter case, the main division is
between optical spectroscopy and mass spectroscopy (or
spectrometry).
Mass spectrometry
generally gives significantly better analytical performance,
but is also significantly more complex. This complexity
translates into higher purchase costs, higher operational costs,
more operator training, and a greater number of components that
can potentially fail. Because optical spectroscopy is
generally less expensive and has performance adequate for
many tasks, it is far more common. Atomic absorption
spectrometers are one of the most commonly sold and used
analytical devices.
Optical Spectroscopy
Electrons exist in energy levels
within an atom. These levels have well defined energies and
electrons moving between these levels must absorb or emit an
energy equal to the difference (or "quantum of energy")
between them. In optical spectroscopy, the energy absorbed to
move an electron to a more energetic level and/or the energy
emitted as the electron moves to a less energetic energy level
is in the form of a photon (a particle of light). Because
this energy is well-defined, an atom's identity (i.e. what
element it is) can be identified by the energy of this
transition. The wavelength of light can be related to its
energy. It is usually easier to measure the wavelength of light
than to directly measure its energy.
Optical spectroscopy can be
divided into absorption, emission, and fluorescence.
In
atomic absorption spectroscopy, light is passed through
a collection of atoms. If the wavelength of the light has energy
corresponding to the energy difference between two energy levels
in the atoms, a portion of the light will be absorbed. The
relationship between the concentration of atoms, the distance
the light travels through the collection of atoms, and the
portion of the light absorbed is given by the
Beer-Lambert law.
The energy stored in the atoms
can be released in a variety of ways. When it is released as
light, this is known as fluorescence.
Atomic fluorescence spectroscopy measures this
emitted light. Fluorescence is generally measured at a 90° angle
from the excitation source to minimize collection of scattered
light from the excitation source, often such a rotation is
provided by a
Pellin-Broca prism on a turntable which will also separate
the light into its spectrum for closer analysis. The
wavelength once again tells you the identity of the
atoms. For low absorbances (and therefore low concentrations)
the intensity of the fluoresced light is directly
proportional to the concentration of atoms. Atomic
fluorescence is generally more sensitive (i.e. it can
detect lower concentrations) than atomic absorption.
Strictly speaking, any
measurement of the emitted light is emission spectroscopy, but
atomic emission spectroscopy usually does not include
fluorescence and rather refers to emission after excitation
by thermal means. In a manner similar to fluorescence, the
wavelength of the emitted photon once again tells you the
identity of the atoms. And again, the intensity of
the emitted light is directly proportional to the concentration
of atoms.
Atomic Mass Spectrometry
Atomic mass spectrometry
is similar to other types of
mass spectrometry in that it consists of an ion source, a
mass analyzer, and a detector. The identities of atoms are
determined by their mass-to-charge ratio (via the mass analyzer)
and their concentrations are determined by the number of ions
detected. Although considerable research has gone into
customizing mass spectrometers for atomic ion sources, it is the
ion source that differs most from other forms of mass
spectrometry. These ion sources must also atomize samples, or an
atomization step must take place before ionization. Atomic ion
sources are generally modifications of atomic optical
spectroscopy atom sources.
III. Mass Spectrometry
Mass spectrometry is used to determine the molecular weight of a
compound, as well as its molecular structure from the fragmentation
pattern. The device measures the mass-to-charge ratio of ions, or
m/e value. Each kind of ion has a particular m/e value.
This measurement is achieved by ionizing the sample and separating ions of differing masses
and recording their relative abundance by measuring relative intensities of ion
flux. Techniques
such as inductively coupled mass spectrometry is used to probes the
masses of individual atoms.
Thus, different compounds have
different atomic masses, and this fact is used in a
mass spectrometer to determine what chemicals are present in
a sample. For example, table salt (NaCl) is vaporized and then
ionized into electrically charged ions (Na+ and Cl-).
The sodium cations (Na+) and chloride anions (Cl-) have specific
atomic weights. They also have an electrical charge, which means
that their path can be controlled with an electric or magnetic
field.
The ions are sent into an
acceleration chamber and passed through a slit in a metal sheet.
A magnetic field is applied to the chamber. The field pushes
each ion perpendicular to the plane defined by the particles
direction of travel and the magnetic field lines. They are then
deflected onto a detector. The lighter ions are deflected more
than the heavier ions because (Newton's
second law of motion) the acceleration of a particle is
inversely proportional to its mass. Likewise, the magnetic field
can push the lighter ions further, thereby giving them a larger
deflection, than the heavier ions. The detector measures exactly
how far each ion has been deflected. From this data, the ion's
'mass-to-charge ratio' can be worked out. Thus, it is possible
to determine with a high level of certainty the chemical
composition of the original sample.
The set of ions is analyzed in
such a way that a signal is obtained for each value of m/e that
is represented by the data. The intensity of each signal
reflects the relative abundance of the ion producing the signal.
The largest peak is called the base peak. Its intensity
is taken as 100, and the intensities of the other peaks are
expressed relative to it. Thus, the results are
'normalized' with respect to the largest peak.
A plot of the relative peak
intensities vs. m/e value is called a mass
spectrum, and is highly characteristic of a particular
compound. Mass spectra can be used to:
1) Establish the identity
of two compounds
2) Help identify the
structure of an new or unknown compound.
Two compounds are shown to be
identical by the fact that they have identical physical
properties (m.p., b.p., density, refractive index, etc.). The
greater the number of physical properties measured, the stronger
the evidence. Now, a single mass spectrum amounts to dozens of
physical properties, since it shows the relative abundance of
dozens of different fragments. Thus, if we measure the
mass spectrum of an unknown compound and find it to be
identical with the spectrum of a previously
reported compound of known structure, then we can conclude that
the two compounds are identical.
The mass spectrum can give a
precise molecular weight, as well as indicating the
presence of certain structural units.
Results
The mass spectrum of
benzene is relatively simple and illustrates some of the information that
mass spectrometry provides. The most intense peak in the mass spectrum is called
the base peak and is assigned a relative intensity of 100. Ion abundances are
proportional to the peak intensities and are reported as intensities relative to
the base peak. The base peak in the mass spectrum of benzene corresponds to the
molecular ion (M+) at m/z = 78.


Benzene does not
undergo extensive fragmentation. None of the fragment ions in its mass spectrum
are as abundant as the molecular ion. The small peak at m/z = 79 is due
to the presence of the isotope C-13 in place of one of the C-12 atoms in the
ring. Indeed, because all hydrocarbon compounds contain carbon and most contain
hydrogen, similar isotopic clusters will appear in the mass spectra of
all organic compounds.
Unlike the case
of benzene, in which ionization involves the loss of a pi electron from the
ring, electron-impact-induced ionization of chlorobenzene involves the loss of
an electron from an unshared pair of electrons. The molecular ion then fragments
by C-Cl bond cleavage.

The peak at m/z
= 77 in the mass spectrum of chlorobenzene is attributed to this fragmentation
and the resulting phenyl cation. The peak at m/z = 77 + 35 = 112 would then
correspond to the molecular ion of chlorobenzene.

Some classes of
compounds so prone to fragmentation that the molecular ion peak is very
weak. The base peak in most unbranched alkanes, for example, is located
at m/z = 43 (corresponding to the propyl cation). This relatively weak base peak
is followed by peaks of decreasing intensity at m/z values of 57, 71, 85, etc.
corresponding to cleavage of each possible C-C bond in the molecule. The
pattern is evident in the mass spectrum of decane (below), and the points of
cleavage are indicated in the diagram below.


Ion /Atom
Sources
Sources can be adapted in many ways, but the
lists below gives the general uses of a number
of sources. Of these, flames are the
most common due to their low cost and their
simplicity. Although significantly less common,
inductively-coupled plasmas, especially
when used with mass spectrometers, are
recognized for their outstanding analytical
high performance and their versatility.
For all
atomic spectroscopy, a sample must be
vaporized and atomized.
Alternatively, for atomic mass
spectrometry, a sample must also
be ionized. Vaporization,
atomization, and ionization are often, but not
always, accomplished with a single source.
Alternatively, one source may be used to
vaporize a sample while another is used to
atomize (and possibly ionize). An example of
this would be laser ablation inductively-coupled
plasma atomic emission spectrometry. A
laser is used to vaporize a solid sample and an
inductively-coupled plasma is used to atomize
the vapor.
With the
exception of flames and graphite furnaces
(commonly used for atomic absorption
spectroscopy) most sources are used
primarily for atomic emission spectroscopy.
Liquid-sampling sources include:
- Flames and
sparks (atom)
-
Inductively-coupled plasma (atom & ion)
- Graphite
furnace (atom)
- Microwave
plasma (atom & ion)
-
Direct-current plasma (atom & ion)
Solid-sampling sources include
- Lasers
(atom & vapor)
- Glow
discharge (atom & ion)
- Arc (atom
& ion)
- Spark
(atom & ion)
- Graphite
furnace (atom & vapor)
Gas-sampling
sources include
- Flame
(atom)
-
Inductively-coupled plasma (atom & ion)
- Microwave
plasma (atom & ion)
-
Direct-current plasma (atom & ion)
- Glow
discharge (atom & ion)
IV. Nuclear Magnetic Resonance
Nuclear magnetic resonance (NMR) spectroscopy is one of the principal
non-destructive techniques used to obtain physical, chemical, electronic and
structural information. This method identifies different nuclei (& elements)
from their magnetic properties, or their magnetic moment. This is the most
important and commonly used spectroscopic technique for organic chemists, often
permitting complete assignment of atom connectivity and even stereochemistry
given the proper set of spectroscopy experiments (e.g.
correlation spectroscopy).
Nuclear magnetic resonance
(NMR) is a physical phenomenon based upon the
quantum mechanical magnetic properties of an atom's nucleus.
All nuclei that contain odd numbers of protons or neutrons have
an intrinsic
magnetic moment and
angular momentum. The most commonly measured nuclei are
hydrogen-1 (the most receptive isotope at natural abundance)
and carbon-13 (CMR), although nuclei from isotopes of
many other elements can also be observed. NMR studies magnetic
nuclei by aligning them with a very powerful external magnetic
field and perturbing this alignment using an electromagnetic
field. The resulting response to the external perturbing
electromagnetic magnetic is the phenomena that is exploited in
nuclear magnetic resonance spectroscopy and
magnetic resonance imaging.
NMR spectroscopy is one of the
principal techniques used to obtain physical, chemical,
electronic and structural information about molecules. It is a
powerful technique that can provide detailed information on the
topology, dynamics and three-dimensional structure of molecules
in solution and the solid state. Also, nuclear magnetic
resonance is one of the techniques that has been used to build
elementary
quantum computers.
V. Crystalline Structure: Diffraction
This is the most precise method for determining
molecular geometry. However, it is very difficult to grow crystals of
sufficient size and high quality to get a clear picture. Thus it remains a
secondary form of analysis. Crystallography has seen especially extensive use in
biochemistry (for protein structure determination) and in the characterization
of organometallic catalysts, which often possess significant symmetry. Primary
methods include
X-ray diffraction,
neutron diffraction and
electron diffraction.
95% of all
solid matter which exists on the surface of the earth has a distinctive and
identifiable crystalline structure. This means there is a certain degree
of long-range order which exists between the individual molecules or
molecular building blocks. Thus the molecules in a single crystal can be thought
of as being composed of a series of parallel planes.
www.ill.fr/dif/3D-crystals/index.html
X-ray diffraction is a
technique in
crystallography in which the pattern produced by the
diffraction of
X-rays through the closely spaced lattice of atoms in a
crystal is recorded and then analyzed to reveal the ordered nature of the
crystal lattice. This generally leads to an understanding of the material and
molecular structure of a substance. The spacing in the crystal lattice can be
determined using
Bragg's law. The electrons that surround the atoms, rather than the atomic
nuclei themselves, are the entities that physically interact with the incoming
X-ray photons.
X-ray
diffraction thus relies on the principle of superposition of waves.
When waves interact with one another, they can do so either destructively or
constructively. In the former case, their incoherent structural features (peaks
and troughs) are not “in sync” with each other, and the result is “noise”
(destructive interference) as opposed to “music”. When the coherent waves are
‘in sync”, their peaks overlap in space to produce a single amplified wave of
identical wavelength. This process is known as
constructive interference.
Multiple
X-rays entering a highly ordered crystalline environment find themselves
reflected from parallel planes whose spacing (or distance of planar separation)
is similar to the wavelength of the incident radiation. This ideal condition
results in constructive interference of the X-rays. The result is the
characteristic periodic arrangement of light and dark bands of
local intensity maxima and minima known as the diffraction pattern of the
X-rays as they are reflected form the planes inside of the crystal. You might
think of these concentric bands as being analogous to the ripples (peaks
and troughs) in a pool as you drop in a stone. All waves behave in
such a similar fashion.

Thus, when a material (sample) is irradiated with a parallel beam of
monochromatic X-rays, the atomic lattice of the sample acts as a three
dimensional diffraction grating causing the X-ray beam to be diffracted to
specific angles. The diffraction pattern includes both the position
(angles) and intensities of the diffracted beam.
Angles are used to calculate the distance d of planar separation
or atomic spacing (d-spacing). The atomic spacing (or lattice parameter) d
is related to the angle of incidence
by
the following relationship, which his known as Bragg’s equation:
n = some whole number integer
=
wavelength of incident radiation
Because
every crystalline material will give a characteristic diffraction pattern
and can act as a unique fingerprint, the position (d) and intensity (I)
information are used to identify the type of material by comparing them
with patterns for over 80,000 experimental data entries. These entries can be
accessed in the International Powder Diffraction File (PDF) database,
complied by the Joint Committee for Powder Diffraction Standards (JCPDS).
By this
method, identification of any crystalline compounds, even in a complex
sample, can be made. The position (d) of diffracted peaks also provides
information about how the atoms are arranged within the crystalline compound (unit
cell size or lattice parameter). The intensity information is used to
assess the type and nature of atoms. Determination of lattice parameter can also
be helpful in understanding the extent of solid solution: complete or
partial substitution of one element for another, as in some metal alloys.
Thus, X-ray diffraction is a versatile, non-destructive technique that
reveals detailed information about the chemical composition and crystallographic
structure of natural and manufactured materials. Areas of application are
quite wide and include metals, organic and inorganic compounds. Also
included are airborne dusts and particulate matter, hazardous inorganic
chemicals, asbestos, metals, bio-materials (bone or dental implants),
pharmaceuticals, catalysts, polymers, ceramics and composites, rocks and
minerals, clays & soils, semiconductors, and corrosion products.
Unfortunately, all glassy (or vitreous) solids are non-crystalline
in nature. Thus their molecular arrangements do not reflect anything but some
limited degree of short-range order, and do not create the type of constructive
interference necessary to produce a recognizable X-ray diffraction pattern. Thus
the use of X-ray diffraction for chemical
identification is limited to crystalline solids only.

Even though X-rays are the most common
choice. For some purposes electrons or neutrons are used, which is possible due
to the wave properties of the particles. These methods are referred to as:
neutron
diffraction and
electron diffraction.
Since neutrons are electrically
neutral (not charged), they do not interact with the electron
cloud surrounding the atom (unlike x-ray and electron
diffraction). The neutrons will only interact with the nucleus
of the atom. Thus, neutron diffraction reveals the atomic
structure only, and not charge distribution around the atom
(although the two are usually very similar).
Neutron diffraction reveals
structural details of the target material, which are measured by
recording the way in which neutrons are deflected. Neutrons can
also change their speed during the scattering experiment. This
is used to study the types of vibrations that can occur in the
solid. One important difference between neutron and X-ray
diffraction is that neutrons are sensitive to magnetic forces in
the material.
Electron diffraction experiments are
usually performed in a
transmission electron microscope (TEM), or a
scanning electron microscope (SEM) as
electron backscatter diffraction. In these instruments, the electrons are
accelerated by an electrostatic potential in order to gain the desired energy
and wavelength before they interact with the sample to be studied.
VI. UV / Vis Spectrophotometry
Spectrophotometry
is more specific than the general term spectroscopy, in that spectrophotometry
deals with visible (white) light and ultraviolet (UV) radiation
as opposed to the entire electromagnetic spectrum.
In between the IR and UV
portions of the electromagnetic spectrum, light of wavelength between about 400
nm and 750 nm is visible to the naked eye. UV spectrometers commonly measure
absorption of light in the visible and near UV region (200 -750 nm). This light
is of higher frequency (and greater energy) than IR light and, when it is
absorbed by a molecule, the changes it produces require greater energy. These
are typically changes in electronic states.
The absorption of light is
typically due electronic transitions in molecules (or
"color centers").
Conjugation /
Color Centers
The UV
spectrum is not used primarily ot show ht presence of individual functional
groups. It is rather used to show relationships which exist between functional
groups.
UV/VIS
spectroscopy is often used to determine the
degree of conjugation in the system:
1)
Conjugation between two or more C=C double (or triple) bonds.
2)
Conjugation between C=C and C=O double bonds.
3)
Conjugation between double bonds and an aromatic ring.
4) The
presence of an aromatic ring.
It can, in addition, reveal the number and location (but not
identity) of substituent groups attached to the carbon atoms of the
conjugated system.
In a
transition to a higher electronic level, a molecule can go from any of a number
of sublevels (vibrational and rotational states) to any of a number of other
sublevels. Thus UV absorption bands are more broad than those seen
in a typical IR spectrum. As opposed to the many sharper peaks of an IR
spectrum, a typical UV spectrum shows only a few broad humps. Parameters
typically used to describe such a spectrum are the position of the top of the
hump (λmax)
and the intensity of that absorption (ε,
the extinction coefficient).
A sigma electron is held
tightly, and a good deal of energy is required to excite it (corresponding to
short wavelength UV radiation). It is often excitations of the comparatively
loosely held n and pi electrons that appear in the longer wavelength
(near UV) spectrum. And, of these, only jumps to the lower (more stable) excited
states.
Conjugation of double bonds lowers the energy required for an electron to shift
from a stable (bonding) orbital to a more unstable (non-bonding) orbital. Thus,
absorption moves to longer wavelengths, where it can be more conveniently
measured. If there are enough double bonds in conjugation, absorption will
move into the visible region, and the compound will be colored.
Beta-carotene is a yellow pigment found in carrots and green leaved plants and
vegetables (such as spinach) and is a precursor of vitamin A. It contains 11 C=C
double bonds in conjugation, and owes its color to absorption at the violet end
of the visible spectrum
λmax
= 451 nm.

Molar
Concentration
An
essential piece of information about any molecular species is how much of it is
present. Quantitative measures of concentration are essential in
all scientific fields. Of all the methods that have been devised for measuring
concentration, by far the most widely applied is absorption spectrophotometry
(especially in liquid samples).
A spectrophotometer (or
photometer) is a device for measuring light intensity. The instrument
typically measures the intensity as a function of the color, frequency or
wavelength of light.
The amount of light that a sample absorbs at a particular
wavelength is measured and used to determine the concentration of the sample by
comparison with appropriate standards or reference data.
The most
useful measure of light absorption is the absorbance (A), also commonly called
the optical density (OD). The absorbance is defined as:

=
Intensity of light incident on the sample
=
Intensity of light transmitted by the sample

The
absorbance of a sample at any given wavelength can be related to the
concentration of the absorbing species through Beer's law:
A = c
ε
c is concentration (usually measured in moles per liter)
ε is
a proportionality constant (the molar extinction coefficient)
The value
of ε is
a function of both the particular compound being measured and the wavelength.
Chlorophylls typically have an
ε value
of about 100,000 liters / mole. When more than one component of a complex
mixture absorbs at a given wavelength, the absorbances due to the individual
components are generally additive. For a review of the mechanisms
responsible for the absorption of light:
www.shu.ac.uk/schools/sci/chem/tutorials/molspec/uvvisab1.htm
The
essential parts of a spectrophotometer include a light source, a wavelength
selection device such as a monochromator or filter, a sample chamber, a light
detector, and a readout device (often include a computer for data storage and
spectral analysis). The most useful machines scan the wavelength of the light
that is incident on the sample and produce, as output, spectra of absorbance
versus wavelength.
The choice
of a source will vary depending on what type of radiation is desired. For
visible radiation, an ordinary tungsten bulb provides a convenient
source. In the UV region, a hydrogen or deuterium discharge lamp is
normally used. For IR radiation, a heated molded rod containing a mixture
of rare earth oxides provides a good source.
The
function of the monochromator is to select a single wavelength or
frequency from the light source. Colored glass filters can provide a convenient
solution to this problem. Higher precision may require a prism or diffraction
grating.
While
glass and quartz are often used for as cell materials, special sample cells
composed of basic salts, such sodium chloride or potassium bromide, are commonly
used because they will not absorb light over a wide range of the IR portion of
the electromagnetic spectrum.
The
detector measures the quantity of radiation that passes thru the sample by
converting it to an electrical signal. UV and visible spectrophotometers employ
photoelectric tube detectors. A signal is generated when the photons strike the
tube surface to produce a current that is directly proportional to the intensity
of the light that is transmitted through the sample. When this signal is
compared to an experimental “blank” (the intensity of light transmitted to he
detector in the absence of an absorbing material), the absorbance of a substance
can be determined at selected frequencies or wavelengths.
The
signal from the detection system is then fed into a recorder, which plots
absorbance as a function of wavelength or frequency. Modern spectrophotometers
are designed to trace an entire absorption spectrum automatically. Many are
equipped with Fourier transform capabilities, which are used to decode
the measured signals of all wavelengths simultaneously and record the generated
data.
An ultraviolet-visible spectrum is
essentially a graph of light absorbance vs. wavelength in
a range of ultraviolet or visible regions. Such a spectrum can
often be produced by a more sophisticated spectrophotometer.
Wavelength is often represented by the symbol λ. Similarly, for
a given substance, a standard graph of extinction coefficient ε
vs. wavelength λ may be made or used if one is already
available. Such a standard graph would be effectively
"concentration-corrected" and thus independent of concentration.
For the given substance, the wavelength at which
maximum absorption in the spectrum occurs is called
λmax (Lambda-max).
Woodward-Fieser rules are a
set of empirical observations which can be used to predict λmax,
the wavelength of the most intense UV/Vis absorption, for
conjugated organic compounds such as dienes and
ketones.
VII. Chromatography
Although
spectrophotometry is exceedingly useful in the chemical identification of many
organic materials, its optimum utilization requires that a material be in a
relatively high state of chemical purity. Because the impurity levels of any
given sample is often beyond the control of the analytical chemist, this
criterion is often not met. For this reason, the analytical technique of
chromatography is widely applied for the analysis of physical evidence.
Chromatography is a means of separating and tentatively identifying the
components of a chemical mixture.
It involves passing a mixture dissolved in a
"mobile phase" through a stationary phase, which separates the
analyte to be measured from other molecules in the mixture and allows it to
be chemically isolated.
For
example, illicit drugs and other controlled substances which are sold on the
street are not manufactured to meet government labeling standards. Instead the
may be “diluted” with practically any material that is readily available to the
dealer in order to increase the quantity and associated street value of the
product. Hence, the task of identifying an illicit drug preparation would be an
arduous one without the aids of chromatographic methods to first separate the
mixture into its components.
The theory
of chromatography has as its basis the observation that chemical substances
have a tendency to partially escape into the surrounding environment when
dissolved in a liquid or when absorbed on a sold surface. In short, the
distribution or “partitioning” of a gas between the liquid and gas phases is
determined by the solubility of the gas in the liquid. The higher
its solubility, the greater the tendency of the gas molecules to remain in the
liquid phase.
More
importantly, if two different gases (analytes A & B) are dissolved in the
same liquid, each will reach a state of chemical equilibrium independently of
one another. Their solubilities will be determined
solely by their respective solubilities in the liquid.
The
essence of chromatography is that some (but not all) substances will actually
have a tendency to move in a specific direction over a stationary (or
fixed) liquid phase in response to an applied flow. This motion is
enhanced by forcing the motion of the air over the liquid.
Interestingly, those components containing the highest number of molecules
in the mobile gas phase (or present at higher number densities) will
move more rapidly in response to the air flow. Eventually, when the entire
gas phase (air + analytes) has advanced a reasonable distance, the component
with the higher mobility will become entirely separated from the others.
The three
different types of chromatography commonly used in the laboratory include:
Gas Chromatography (GC), High-Performance Liquid Chromatography (HPLC), and
Thin Layer Chromatography (TLC). We briefly outline those here.
Gas Chromatography (GC)
Gas
chromatography (or
gas-liquid chromatography) separates mixtures on the basis of their
distribution between a stationary liquid phase and a mobile gas phase. A
chemically inert carrier gas (e.g. nitrogen or helium) flows through a column
made of stainless steel or glass. The liquid sample under investigation is
injected into a heated injection port by syringe, where it is immediately
vaporized and swept into the column by the carrier gas. As the carrier gas flows
through the column, it carries along with it the components of a mixture that
have been injected into the column. Components will be separated according to
their relative mobilities (and corresponding solubilities) in the mobile gas
phase.
As each
component emerges from the column, it enters a detector. One type of detector
uses a flame to ionize the emerging chemical substance, thus generating an
electrical signal. The signal is recorded onto a strip-chart recorder as a
function of time. This record is referred to as a
chromatogram.
A typical
gas chromatogram will show a series of peaks, each one corresponding to a
specific component of the chemical mixture. The time required for a component to
emerge form the column from the time of its injection is known as the
retention time. This experimentally determined parameter serves as a useful
identifying characteristic of a material (tho it may vary for different
instrument configurations).
Gas
chromatography has an added advantage in that it is extremely sensitive
and can yield quantitative results. The amount of substance passing thru
the GC detector is directly proportional to the area underneath the recorded
peak. The amount of sample can often be determined to within the nearest
billionth of a gram.
One
important extension of this application in forensic science is the technique of
Pyrolysis GC. Many solid materials commonly encountered as physical
evidence (paint chips, fibers, plastics, etc.) cannot be readily dissolved in a
solvent for liquid injection. However, materials such as these can be
heated or pyrolyzed to high temperatures (500 – 1000 degrees C) so that they
will decompose into numerous gaseous products. Pyrolyzers have been designed in
order to permit these gaseous products to enter the carrier gas stream, flowing
into and thru the GC column.
Liquid Chromatography (HPLC)
In the
case of high–performance liquid phase chromatography (HPLC),
the stationary phase is a finely ground particulate solid. The mobile (carrier)
phase is a liquid that is pumped thru a column filled with fine solid particles
whose surfaces have been chemically treated. The major advantage of HPLC is that
the entire process takes place at room temperature. Examples of materials
where HPLC may be the preferred method of choice may include organic
explosives, which are generally heat sensitive. Also included are heat
sensitive psycho-active drugs, such as LSD.
Thin-Layer Chromatography
The
technique of thin-layer chromatography (TLC)
incorporates a solid stationary phase and a mobile liquid phase to effect the
separation of the constituents of a mixture. A thin-layer plate is
prepared by coating a glass plate with a thin film of granular material,
which is bonded to the plate with plaster of Paris.
Solid
samples must first be dissolved and a few microliters of the solution spotted
with a capillary tube or syringe onto the granular surface near the lower edge
of the plate. A liquid sample may be applied directly to the plate in a similar
manner. The plate is then placed upright into a closed chamber that contains a
selected liquid, with care that the liquid does not touch the sample spot.
The liquid
will slowly begin to rise up along the surface of the plate by capillary
action. The rising liquid serves as the mobile carrier phase. As it
moves past the sample spot, the components of the sample become distributed
between the stationary solid phase and the mobile liquid phase. Those components
with a greater solubility in the mobile liquid phase will travel up the surface
of the plate at a faster speed. When the liquid front has moved a sufficient
distance (usually 10 cm) the development is complete, and the plate is removed
from the chamber and dried.
Because
most components are colorless (exceptions: organic pigments such as chlorophyll
and beta-carotene) no separation will be noticeable after development unless the
materials are visualized. To accomplish this, the plates are placed under UV
light, revealing those materials that fluoresce as bright spots on a
dark background. For identification purposes, a high confidence level
necessitates the use of experimental standards which can be compared to the
results for confirmation of the suspect material’s chemical composition.
Migration distances
must therefore be matched with similar experiments on similar compounds
using similar experimental configurations and apparatuses. If the distances are
identical, then a tentative identification can be made. However, it must be
cautioned that such identification cannot be considered definitive for the
possibility always exists that numerous other substances can migrate the same
distance up the plate when chromatographed under similar conditions. Thus,
thin-layer chromatography cannot by itself provide an absolute and undisputed
identification. Thus, TLC must be used in conjunction with other testing
procedures in order to provide absolute proof of chemical identity.
GC / MS
Many of
the uncertainties associated with vapor phase chromatography (GC) have
been overcome by linking the gas chromatograph to a mass spectrometer to
yield an extremely powerful combination known as gas chromatography / mass
spectrometry (GC / MS).
The
separation of a mixture’s components is first accomplished on the gas
chromatograph. A direct connection between the GC column and the mass
spectrometer then allows each component to flow into the spectrometer as it
emerges form the gas chromatograph.
In the
mass spectrometer, the material enters a high-vacuum chamber where a beam of
high-energy electrons is aimed at the sample molecules. The electrons collide
with the molecules, causing them to lose electrons and to acquire a positive
charge. These positively charged (+) cations are highly unstable and rapidly
decompose into smaller fragments. These fragments then pass through an electric
or magnetic filed, where they are separated according to their masses. The
advantage of this technique is that no two substances produce the same
fragmentation pattern.
Hence,
this fragmentation pattern serves as a “fingerprint” of a specific
chemical substance. It is also sensitive to minute concentrations. With the data
obtained from a GC/ MS determination, a forensic analyst can, with one
instrument, separate the components of a complex drug mixture and then
unequivocally identify each substance present in the mixture.